How to calibrate an oscilloscope for phasing
Calibrating X-Y Phasing
X input
Depending on the type of oscilloscope, the X input will be applied either via the External Trigger connector, or via Channel 1, with suitable switching. In either case, the same signal of 50 mV, 50 kHz will be applied to both X and Y inputs.
a. Initial setup
The oscilloscope controls are calibration set as follows:
- Vertical mode: X-Y
- Sensitivity: 5 mV/div, both channels
- Ch 1 or X: AC coupled
- Ch 2 or Y: Grounded
- Vertical mode: X-Y
- Vertical position: Central
- Horizontal position: Central
Note: In X-Y phasing tests on a DSO, the highest possible sampling rate is employed. However, the display of captured Lissajous figures may show interrupted segments due to the storage length limitation, until the frequency of the test signal is sufficiently high to capture a complete cycle.
b. Trace acquisition
The display intensity is adjusted until a horizontal trace is just visible (should be 10 divisions long). After the X and Y position controls have been used to center the trace, the intensity and focus controls are adjusted for best display.
c. Phasing check
The common input signal is reduced until the trace is 8 divisions long. Channel 2 (or Y) input mode is switched to DC, and the X and Y position controls are used to center the (now sloping) trace. If the X and Y channels do not introduce any phase error, then the center of the trace will pass through the origin. Phase error between X and Y channels will cause the sloping trace to split into an ellipse, which for small phase errors will be apparent only close to the origin. The trace separation at the origin, along the center horizontal graticule line, should be not greater than 0.4 division for a commonly specified phase shift of 3 degrees.
Return to previous articles
- Início
- Produtos
- Novos Produtos
- Calibração elétrica
- Calibração de RF
- Equipamento de aquisição de dados e de teste
- Calibração de temperatura
- Calibração de umidade
- Calibração de pressão
- Medidores de pistão
- Controlador/Calibradores
- Monitores de pressão
- Balanças de peso morto
- Calibradores manuais e monitores
- Handheld Pressure Calibrators
- Calibração de dados de aéreos
- Monitores ambientais
- Acessórios de calibração de pressão
- Sistemas personalizados de calibração de pressão
- Software de calibração de pressão
- Calibração de fluxo
- Ferramentas de Calibração de processo
- Software de calibração
- Serviço e suporte
- Todos os instrumentos de calibração
- Onde Comprar
- Notícias
- Treinamento e eventos
- Literatura e educação
- Serviço e Suporte
- Service Request (RMA)
- Planos de Serviço
- Suporte técnico
- Base de Conhecimento
- Acreditações
- Centros de assistência autorizados
- Calibration Certificates
- Community Forum
- My MET/SUPPORT
- Manuais de produto (Guias de usuário)
- Folhas de Dados de Segurança (SDS)
- Safety, Service, and Product Notices
- Downloads de software
- Sobre Nós